Turbimax CUE21 / CUE22
Operation
Hauser
19
8
Operation
8.1
Configuring the measuring device
8.1.1
Selecting the output
Function
Options
Info
Output (O/P)
AUTO
CAL
CONFIG
4-20
O/P
Å
• 4-20
• 485
• OFF
4-20
4 mA value
0 to 1000 NTU
Set the lower turbidity limit value (LOLM) that
corresponds to the 4 mA output level.
20 mA value
0 to 1000 NTU
Set the upper turbidity value (UPLM) that
corresponds to the 20 mA output level.
485
Baud
• 1200
• 2400
• 4800
• 9600
• 19200
Set the correct baud rate for operation of the I/O
port.
Addr
1 ... 255
Select the desired device address.
MBUS
• ASCII
• RTU
Select ASCII or RTU.
8.1.2
Configuring the alarms
Function
Options
Info
Alarm 1 (ALM1)
AUTO
CAL
CONFIG
HI
ALM1
Å
• HI
• LO
• OFF
Set point (S/P)
0 to 1000 NTU
Set the desired alarm set point.
Delay on (DLY
▴
)
1 to 30 s
Set the desired number of seconds for the "Delay
on" function.
Delay off (DLY
▾
)
1 to 30 s
Set the desired number of seconds for the "Delay
off" function.
Alarm 2 (ALM2)
• HI
• LO
• OFF
Set point (S/P)
0 to 1000 NTU
Set the desired alarm set point.