LCD backlight brightness
Function
Options
Info
LCD backlight brightness (BRT)
AUTO
CAL
CONFIG
8
Å
BRT
1 to 10
Factory setting: 8
Adjust the backlighting for the display if necessary.
Units
Function
Options
Info
Units (UNIT)
AUTO
CAL
CONFIG
NTU
Å
UNIT
• NTU
• FNU
Factory setting: NTU
Select the desired unit for displaying the turbidity
level: NTU (Nephelometric Turbidity Units) or FNU
(Formazin Nephelometric Units).
Ultrasonic cleaning
The "Ultrasonic cleaning" option is used for continuous cleaning of the flow-through
cuvette. It is not intended to clean cuvettes that are already dirty, or to replace manual
cleaning entirely. However, this system will increase the cleaning intervals dramatically.
To operate correctly, the use of a cuvette with an ultrasonic transducer is required. The
system transmits an ultrasonic frequency through a spring to a piezo transducer that is
joined to the base of the flow-through cuvette. The cuvette's detection system works only
in AUTO mode.
The system can detect the following error conditions:
• Incorrect cuvette installed
• Error has occurred in the transducer
• Transducer not making contact with the spring
If an error has occurred, the message "CLN" appears in the lower part of the display. The
alarms are set and 2mA is sent to the current output. The ultrasonic cleaning function is
switched on as standard. After a cuvette has been inserted, the message "Dry" appears for a
duration of 30 minutes in the lower part of the display. During this time, the desiccant
removes residual moisture from the ultrasonic transducer and no ultrasonic cleaning takes
place. At the same time, the error message "CLN" is reset and appears only after the 30
minutes have elapsed.
Function
Options
Info
Ultrasonic cleaning (CLN)
AUTO
CAL
CONFIG
On
Å
CLN
• On
• OFF
On
The ultrasonic cleaning function can be turned off
if desired.