Diagnostics and troubleshooting
Turbimax CUE21 / CUE22
28
Hauser
Diagnostic
message
Possible cause
Tests or remedial action
FLOW
Sample flow has stopped
Restore flow
CLN
Ultrasonic transducer in flow-
through cuvette not making contact
Rotate the flow-through cuvette slightly to improve
the connection. If the problem persists, replace the
cuvette.
Flow-through cuvette removed
Insert flow-through cuvette
FAIL
Major system fault
Contact sales office.
DRY
Drying period
Disappears automatically after a period of 30
minutes.