Turbimax CUE21 / CUE22
Installation
Hauser
11
A0025699
5
Recommended plumbing
1
Connection for intake tube
7
Connection for drain tube
2
Shutoff clamp
8
Vent for drain tube
3
Connection for intake tube
9
Connection for drain tube
4
Connection for intake tube
10
Use
5
Connection for drain tube
11
Nut
6
Flow controller
12
Tubing (not supplied)
The device is equipped for plumbing using tubing with OD 8 mm (0.31 inch),
ID 4.75 mm (3/16 inch). If the device is going to be exposed to sunlight, opaque
tubing should be used to prevent algae growth.
Plumbing the device
1. Push the shutoff clamp (item 2) over one of the supplied tubes.
2. Connect the tube to the upper part of the intake tubing connection (item 3) and to
the connection on top of the device (item 1).
3. Push the flow controller (item 6) over the second tube supplied.
4. Connect the tube to the upper part of the drain tubing connection (item 7) and to the
connection on top of the device (item 5).
5. Connect a tube with an internal diameter of 4.75 mm (3/16 inch) to the lower part
of the intake tubing connection (item 4) to supply the sensor with a reliable sample
flow.
6. Connect a tube with an internal diameter of 4.75 mm (3/16 inch) to the lower part
of the drain tubing connection (item 9) to route the sensor drain to a suitable drain
on site.
The vent in the drain tube (item 8) allows for pressure compensation, thus
helping to prevent bubble formation in the cuvette.
7. If your CUE21 / CUE22 is integrated in a high-pressure system,
insert the 6:32 seal screw provided into the vent hole and tighten it.