Smartec S CLD132/134 PROFIBUS-PA/-DP
Technical data
Hauser
39
9
Technical data
9.1
Output PROFIBUS-PA
Output signal
PROFIBUS PA acc. to EN 50170 vol. 2, profile version 3.0
PA function
Slave
Transmission rate
31.25 kbps
Signal coding
Manchester II
Response time slave
Approx. 20 ms
Signal on alarm
Status and alarm messages in accordance with PROFIBUS PA, profile version 3.0
Display: error code
Physical layer
IEC 61158-2, MBP (Manchester bus coded)
Bus voltage
9 to 32 V
Bus error consumption
10 mA ±1 mA
Error current consumption
I
FDE
0 mA
9.2
Output PROFIBUS-DP
Output signal
PROFIBUS DP acc. to EN 50170 vol. 2, profile version 3.0
DP function
Slave
Transmission rate
9.6 kbps, 19.2 kbps, 45.45 kbps, 93.75 kbps, 187.5 kbps,
500 kbps, 1.5 Mbps
Signal coding
NRZ code
Response time slave
Approx. 20 ms
Signal on alarm
Status and alarm messages in accordance with PROFIBUS DP, profile version 3.0
Display: error code
Physical layer
RS 485