Smartec S CLD132/134 PROFIBUS-PA/-DP
System integration
Hauser
25
Analyzer Transducer Block
There are two Analyzer Transducer Blocks in the SmarTec S. These are distributed to slots 1
and 2 in the following order:
1.
Main measured value (Main Process Value)
2.
Measured temperature value (Temperature)
Parameter
E+H
Matrix
Slot
Index Size
(bytes)
Type
Acc. Store
Standard parameter
BLOCK_OBJECT
1 - 2
100
20
DS-32*
r
C
ST_REV
1 - 2
101
2
Unsigned16
r
N
TAG_DESC
1 - 2
102
32
Octetstring
r, w
S
STRATEGY
1 - 2
103
2
Unsigned16
r, w
S
ALERT_KEY
1 - 2
104
1
Unsigned8
r, w
S
TARGET_MODE
1 - 2
105
1
Unsigned8
r, w
S
MODE_BLK
Actual
Permitted
Normal
1 - 2
1016
3
DS-37*
Unsigned8
Unsigned8
Unsigned8
r
N
Cst
Cst
ALARM_SUM
1 - 2
107
8
DS-42*
r
D
Block parameter
COMPONENT_NAME
1 - 2
108
32
Octetstring
r, w
S
PV
1 - 2
109
12
DS-60*
r
D
PV_UNIT
1 - 2
110
2
Unsigned16
r, w
S
PV_UNIT_TEXT
1 - 2
111
8
Visible string
r, w
S
ACTIVE_RANGE
1 - 2
112
1
Unsigned8
1: Range 1
r, w
S
AUTORANGE_ON
1 - 2
113
1
Boolean
r, w
S
SAMPLING_RATE
1 - 2
114
4
Time_difference
r, w
S
Gap reserved PNO
1 - 2
115 -
124
NUMBER_OF_RANGES
1 - 2
125
1
Unsigned8
r
N
RANGE_1
1 - 2
126
8
DS-61*
r, w
N