Smartec S CLD132/134 PROFIBUS-PA/-DP
System integration
Hauser
23
Slot/index tables
The device parameters are listed in the following tables. You can access these parameters by
means of the slot and index number.
Each individual block comprises standard parameters, block parameters and manufacturer-
specific parameters to an extent.
Device management
Physical Block
Parameter
E+H
Matrix
Slot
Index Size
(bytes)
Type
Acc. Store
DIR_OBJECT HEADER
1
0
12
Array of unsigned16
r
Cst.
COMP_LIST_DIR_ENTRIES
1
1
32
Array of unsigned16
r
Cst.
COMP_DIR_ENTRIES_CONTINUES
1
2
12
Array of unsigned16
r
Cst.
Parameter
E+H
Matrix
Slot
Index Size
(bytes)
Type
Acc. Store
Standard parameter
BLOCK_OBJECT
1
160
20
DS-32*
r
C
ST_REV
1
161
2
Unsigned16
r
N
TAG_DESC
VAH0
1
162
32
Octetstring
r, w
S
STRATEGY
1
163
2
Unsigned16
r, w
S
ALERT_KEY
1
164
1
Unsigned8
r, w
S
TARGET_MODE
1
165
1
Unsigned8
r, w
S
MODE_BLK
Actual
Permitted
Normal
1
166
3
DS-37*
Unsigned8
Unsigned8
Unsigned8
r
N
Cst
Cst
ALARM_SUM
1
167
8
DS-42*
r
D
Blockparameter
SOFTWARE_REVISION
1
168
16
Visible string
r
Cst
HARDWARE_REVISION
1
169
16
Visible string
r
Cst
DEVICE_MAN_ID
1
170
2
Unsigned16
r
Cst
DEVICE_ID
1
171
16
Visible string
r
Cst
DEVICE_SER_NUM
1
172
16
Visible string
r
Cst
DIAGNOSIS
1
173
4
Octetstring
r
D
DIAGNOSIS_EXTENSION
1
174
6
Octetstring
r
D
DIAGNOSIS_MASK
1
175
4
Octetstring
r
Cst
DIAGNOSIS_MASK_EXTENSION
1
176
6
Octetstring
r
Cst
DEVICE_CERTIFICATION
1
177
32
Visible string
r
N
WRITE_LOCKING 1
178
2
Unsigned16
0: acyclic refused
2457: writeable
r, w
N
FACTORY_RESET
1
179
2
Unsigned16
0x8000:
reset calibration data
0x8001:
reset set data
0x0001:
PNO defaults
all data
2506: warm start
2712: reset bus adr.
r, w
S