Smartec S CLD132/134 PROFIBUS-PA/-DP
System integration
Hauser
27
Manufacturer-specific parameters Smartec S
Parameter
E+H
Matrix
Slot
Index Size
(bytes)
Type
Acc. Store
Measured value
V0H0
3
100
4
Float
r
D
Temperature
V0H1
3
101
4
Float
r
D
Operating mode
V0H2
3
102
1
Unsigned8
0: Conductivity
1: Concentration
r
D
Unit of measurement (concentration) V0H3
3
103
1
Unsigned8
57: %
139: ppm
245: mg/l
106: tds
251: without
r, w
N
Number of decimal places
V0H4
3
104
1
Unsigned8
0: X.xxx
1: XX.xx
2: XXX.x
3: XXXX
r, w
N
Unit of measurement (conductivity)
V0H5
3
105
1
Unsigned8
66: mS/cm
67:
μ
m/cm
240: S/m
r, w
N
Signal damping
V0H6
3
106
1
Unsigned8
r, w
N
Raw value
V0H7
3
107
4
Float
r
D
Current measuring range
V0H9
3
108
1
Unsigned8
r, w
N
Temperature measurement
V1H0
3
109
1
Unsigned8
0: fixed
1: Pt 100
2: Pt 1000
3: NTC
r, w
N
Process temperature
V1H3
3
110
4
Float
r, w
N
Cell constant
V1H4
3
111
4
Float
r, w
N
Installation factor
V1H6
3
112
4
Float
r, w
N
Calibration temperature
V1H8
3
113
4
Float
r, w
N
Temperature correction
V1H9
3
114
4
Float
r, w
N
Contact function
V3H0
3
115
1
Unsigned8
0: Alarm function
1: Limit function
2: Limit + alarm fct.
r, w
N
Switch-on delay
V3H3
3
116
2
Unsigned16
r, w
N
Switch-off delay
V3H4
3
117
2
Unsigned16
r, w
N
Number of binary inputs
V4H0
3
118
1
Unsigned8
r, w
N
Source of binary inputs
V4H1
3
119
1
Unsigned8
0: binary contacts
1: cyclic data
r, w
N
Processed measuring range
V4H2
3
120
1
Unsigned8
r, w
N
Operating mode of processed
measuring range
V4H3
3
121
1
Unsigned8
0: Conductivity
1: Concentration
r, w
N
Substance selection of processed
measuring range
V4H4
3
122
1
Unsigned8
0: NaOH
1: H2SO4
2: H3PO4
3: HNO3
4: User 1...
r, w
N