Smartec S CLD132/134 PROFIBUS-PA/-DP
Diagnostics and troubleshooting
Hauser
35
7
Diagnostics and troubleshooting
System error messages
The DIAGNOSIS and DIAGNOSIS_EXTENSION parameters are generated from the device-
specific errors.
Smartec S PROFIBUS diagnosis
NAMUR
class
Error
no.
Description
DIAGNOSIS
DIAGNOSIS_
EXTENSION
Measured value status
Quality
Sub-Status
Hex*
Failure
E001
Memory error
01 00 00 80 - DIA_HW_ELECTR
01 00 00 00 00
00
BAD
device failure
0C
Failure
E002
Data error in EEPROM
10 00 00 80 - DIA_MEM_CHKSUM
02 00 00 00 00
00
BAD
device failure
0C
Failure
E003
Invalid configuration
00 04 00 80 - DIA_CONF_INVAL
04 00 00 00 00
00
BAD
device failure
0C
Failure
E007
Transmitter faulty
20 00 00 80 - DIA_MEASUREMENT
08 00 00 00 00
00
BAD
device failure
0C
Failure
E008
Sensor or sensor
connection faulty
20 00 00 80 - DIA_MEASUREMENT
10 00 00 00 00
00
BAD
sensor failure
10
Failure
E010
Temperature sensor
defective
20 00 00 80 - DIA_MEASUREMENT
20 00 00 00 00
00
BAD
sensor failure
10
Failure
E025
Limit value for airset offset
exceeded
20 00 00 80 - DIA_MEASUREMENT
40 00 00 00 00
00
BAD
configuration
error
04
Failure
E036
Sensor calibration range
exceeded
20 00 00 80 - DIA_MEASUREMENT
80 00 00 00 00
00
BAD
configuration
error
04
Failure
E037
Sensor calibration range
undershot
20 00 00 80 - DIA_MEASUREMENT
00 01 00 00 00
00
BAD
configuration
error
04
Failure
E045
Calibration aborted
20 00 00 80 - DIA_MEASUREMENT
00 02 00 00 00
00
BAD
configuration
error
04
Failure
E049
Installation factor
exceeded
20 00 00 80 - DIA_MEASUREMENT
00 04 00 00 00
00
BAD
configuration
error
04
Failure
E050
Installation factor
undershot
00 20 00 80 - DIA_MAINTENANCE
00 08 00 00 00
00
BAD
configuration
error
5C
Failure
E055
Measuring range of the
main parameter undershot
20 00 00 80 - DIA_MEASUREMENT
00 10 00 00 00
00
UNCERTAIN
sensor
conversion not
accurate
50
Failure
E057
Measuring range of the
main parameter exceeded
20 00 00 80 - DIA_MEASUREMENT
00 20 00 00 00
00
UNCERTAIN
sensor
conversion not
accurate
50
Failure
E059
Temperature range
undershot
20 00 00 80 - DIA_MEASUREMENT
00 40 00 00 00
00
UNCERTAIN
sensor
conversion not
accurate
50
Failure
E061
Temperature range
exceeded
20 00 00 80 - DIA_MEASUREMENT
00 80 00 00 00
00
UNCERTAIN
sensor
conversion not
accurate
50
Failure
E067
Limit contactor setpoint
exceeded
00 20 00 80 - DIA_MAINTENANCE
00 00 00 04 00
00
UNCERTAIN
non-specific
40
Failure
E077
Temperature outside of the
value table
00 04 00 80 - DIA_CONF_INVAL
00 00 01 00 00
00
BAD
configuration
error
04
Failure
E078
Temperature outside of the
concentration table
00 04 00 80 - DIA_CONF_INVAL
00 00 02 00 00
00
BAD
configuration
error
04
Failure
E079
Conductivity outside of the
concentration table
0 04 00 80 - DIA_CONF_INVAL
00 00 04 00 00
00
BAD
configuration
error
04
Funct.
check
E101
Service function active
–
–