Proline t-mass B 150
Installation
Hauser
13
230
220
210
200
190
180
9
8
7
A
B
C2
C1
A
B
C2
C1
230
220
210
200
190
180
9
8
7
230
220
210
200
190
180
9
8
7
A0015768
A
Internal pipe diameter DN (circular pipe) or internal dimension (rectangular duct)
B
Thickness of pipe wall or of duct wall
C1 Length of mounting set
C2 Length of sensor compression fitting
Determining C1 and C2 (Hauser original parts only)
DK6MB-BXA mounting boss G1A
C1 + C2 = 99 mm (3.90 in)
DK6MB-DXA mounting boss G3/4A
C1 + C2 = 99 mm (3.90 in)
DK6MB-AXA mounting boss 1 NPT
C1 + C2 = 107 mm (4.21 in)
DK6MB-CXA mounting boss 3/4 NPT
C1 + C2 = 102 mm (4.02 in)
Determining C1 and C2 (not limited to Hauser original parts)
C1
Length of pipe connection used
C2 (compression fitting with G1A thread)
39 mm (1.54 in)
C2 (compression fitting with G3/4A thread)
39 mm (1.54 in)
C2 (compression fitting with 1 NPT thread)
47 mm (1.85 in)
C2 (compression fitting with 3/4 NPT thread)
42 mm (1.65 in)
►
Calculating insertion depth
(0.3 · A) + B + (C1 + C2)