Installation
Proline t-mass B 150
16
Hauser
Avoid direct sunlight, particularly in warm climatic regions.
System pressure
Sensor
Depending on the version, please note the details on the name plate .
Max. 20 bar g (290 psi g)
Thermal insulation
A0015763
6.2
Mounting the measuring device
6.2.1
Required tools
For transmitter
For turning the transmitter housing (in increments of 90
°
): Allen screw 4 mm (0.15 in)
For sensor
For the sensor gland: Corresponding mounting tools
6.2.2
Preparing the measuring device
1. Remove all remaining transport packaging.
2. Remove the protective cap from the sensor.
3. Remove stick-on label on the electronics compartment cover.
6.2.3
Mounting the measuring device
WARNING
Danger due to improper process sealing
►
Ensure that the gaskets are clean and undamaged (G 1 A, G ¾ A).
►
Ensure that the correct sealing material has been used (e.g. Teflon tape for NPT 1 , NPT ¾ ).
►
Install the gaskets correctly.