Proline Promass A 100
Electrical connection
Hauser
35
Setting the address
OFF ON
1
2
3
4
5
6
7
8
9
10
1
2
4
8
16
32
64
- Write protection
- Not used
PROFIBUS
address
- Software addressing
A0021265
10
Addressing using DIP switches on the I/O electronics module
1. Disable software addressing via DIP switch 8 (OFF).
2. Set the desired device address via the corresponding DIP switches 1 to 7.
Example (→ 10, 35): 1 + 16 + 32 = device address 49
The device demands rebooting after 10 s. After rebooting, hardware
addressing is enabled with the configured IP address.
7.3.2
Enabling the terminating resistor
PROFIBUS DP
To avoid incorrect communication transmission caused by impedance mismatch,
terminate the PROFIBUS DP cable correctly at the start and end of the bus segment.
• If the device is operated with a baud rate of 1.5 MBaud and under:
For the last transmitter on the bus, terminate via DIP switch 2 (bus termination) and
DIP switch 1 and 3 (Bus polarisation). Setting: ON – ON – ON (→ 11, 36).
• For baud rates > 1.5 MBaud:
Due to the capacitance load of the user and the line reflections generated as a result,
ensure that an external bus terminator is used.
It is generally advisable to use an external bus terminator as the entire segment
can fail if a device that is terminated internally is defective.