Proline Promass A 100
Operation options
Hauser
47
5
3
1
2
4
A0016961
1
Ethernet network
2
Automation system, e.g. "RSLogix" (Rockwell Automation)
3
Workstation for measuring device operation: with Add-on Profile Level 3 for "RSLogix 5000"
(Rockwell Automation) or with Electronic Data Sheet (EDS)
4
Computer with Web browser (e.g. Internet Explorer) for accessing the integrated device Web
server or with "FieldCare" operating tool with COM DTM "CDI Communication TCP/IP"
5
Ethernet switch
8.3.4
Establishing a connection
For device version with EtherNet/IP communication type
Via service interface (CDI-RJ45) and "FieldCare" operating tool
1. Start FieldCare and launch the project.
2. In the network: Add a device.
The
Add device
window opens.
3. Select the
CDI Communication TCP/IP
option from the list and press
OK
to
confirm.
4. Right-click
CDI Communication TCP/IP
and select the
Add device
option in the
context menu that opens.
5. Select the desired device from the list and press
OK
to confirm.
The
CDI Communication TCP/IP (Configuration)
window opens.
6. Enter the device address in the
IP address
field and press
Enter
to confirm:
192.168.1.212 (factory setting); if the IP address is not known, set DIP switch
No. 10 on the I/O electronics module to ON, restart the device and enter the
default setting for the IP address 192.168.1.212.
7. Establish the online connection to the device.