Proline Promag L 800
Electrical connection
Hauser
45
7.6 Guaranteeing the degree of protection of the measuring device
"
Caution!
Do not loosen the threaded fasteners of the sensor housing, as otherwise the degree of
protection guaranteed by Hauser no longer applies.
To guarantee the degree of protection of the measuring device.
Perform the following steps after electrical connection:
• Check that the housing seals of the connection and electronics compartment are clean and
fitted correctly. Dry, clean or replace the seals if necessary.
• Tighten all housing screws and screw covers.
• Firmly tighten the cable glands.
• To ensure that moisture does not enter the cable entry, route the cable so that it loops down
before the cable entry ("water trap")
• Insert dummy plugs into unused cable entries.
A0013960
Fig. 29: Cable looped down before the cable entry
7.7 Post-connection check
Is the measuring device undamaged (visual inspection)?
Do the cables used comply with the requirements?
Do the cables have adequate strain relief?
Are all cable glands installed, firmly tightened and correctly sealed? Cables looped as “water traps”?
Does the supply voltage match the specifications on the transmitter nameplate?
Is the ribbon cable for the display and operating module correctly routed in the housing?
Is the terminal assignment correct?
Have the batteries been inserted and secured correctly?
Is the DIP switch set to the correct setting?
If power is supplied, is the measuring device ready for operation (the red LED is flashing) and does information
appear on the display if an operating key is pressed for >1 second?
Are all housing covers installed and tightened with the correct tightening torque?