7.1.5 Preparing the measuring device
• Remove any dummy plugs.
"
Caution!
Poor sealing in the housing can affect the operational reliability of the measuring device. Use
suitable cable glands that correspond to the degree of protection.
If the measuring device is supplied without cable glands, provide appropriate cable glands for
the connecting cable that comply with IP protection requirements.
• If the measuring device is supplied with cable glands, observe the cable specifications.
Cable entry
Compact version
Remote version
A0016457
Fig. 18: Cable entries for the compact version
1
Connection terminal for GSM antenna (optional)
2
External power supply (optional)
3
Inputs/outputs
A0016458
Fig. 19: Cable entries for the remote version
1
Connection terminal for GSM antenna (optional)
2
External power supply (optional)
3
Inputs/outputs
4
Coil current cable
5
Electrode cable
1
2
3
1
2
3
4
5