Proline Promag 55S
Hauser
21
Conductivity
Minimum conductivity:
•
≥
5
μ
S/cm for all liquids (incl. demineralized water)
!
Note!
In the remote version, the required minimum conductivity is also influenced by the length of the cable
→
ä
18.
Limiting medium pressure
range (nominal pressure)
• EN 1092-1 (DIN 2501): PN 10 (DN 200 to 600 / 8 to 24"), PN 16 (DN 65 to 600 / 2½ to 24"),
PN 25 (DN 200 to 600 / 8 to 24"), PN 40 (DN 15 to 150 / ½ to 6")
• ANSI B16.5: Class 150 (DN ½ to 24"), Class 300 (DN ½ to 6")
• JIS B2220: 10K (DN 50 to 300 / 2 to 12"), 20K (DN 15 to 300 / ½ to 12")
• AS 2129: Table E (DN 25 / 1", DN 50 / 2")
• AS 4087: Cl. 14 (DN 50 / 2")
Pressure tightness (lining)
Pressure tightness in SI units [mbar]
Promag S
Nominal
diameter
Measuring tube
lining
Resistance of measuring tube lining to partial vacuum (SI units)
Limit values for abs. pressure [mbar] at various fluid temperatures
[mm]
25 °C
50 °C
80 °C
100 °C
130 °C
150 °C
180 °C
25 to 600
Polyurethane
0
0
-
-
-
-
-
65 to 600
Natural rubber
0
0
-
-
-
-
-
65 to 600
Hard rubber
0
0
0
-
-
-
-
Promag S
Nominal
diameter
Measuring tube
lining
Resistance of measuring tube lining to partial vacuum (SI units)
Limit values for abs. pressure [mbar] at various fluid temperatures
[mm]
25 °C
80 °C
100 °C
130 °C
150 °C
180 °C
15
PTFE
0
0
0
100
–
–
25
PTFE / PFA
0/0
0/0
0/0
100/0
–/0
–/0
32
PTFE / PFA
0/0
0/0
0/0
100/0
–/0
–/0
40
PTFE / PFA
0/0
0/0
0/0
100/0
–/0
–/0
50
PTFE / PFA
0/0
0/0
0/0
100/0
–/0
–/0
65
PTFE / PFA
0/0
*
40/0
130/0
–/0
–/0
80
PTFE / PFA
0/0
*
40/0
130/0
–/0
–/0
100
PTFE / PFA
0/0
*
135/0
170/0
–/0
–/0
125
PTFE / PFA
135/0
*
240/0
385/0
–/0
–/0
150
PTFE / PFA
135/0
*
240/0
385/0
–/0
–/0
200
PTFE / PFA
200/0
*
290/0
410/0
–/0
–/0
250
PTFE
330
*
400
530
–
–
300
PTFE
400
*
500
630
–
–
350
PTFE
470
*
600
730
–
–
400
PTFE
540
*
670
800
–
–
450
PTFE
Partial vacuum is impermissible!
500
PTFE
600
PTFE
* No value can be quoted.
Содержание Proline Promag 55S
Страница 47: ...Proline Promag 55S Endress Hauser 47...