Operating conditions: Installation
Installation instructions
Location
The accumulation of air or gas bubbles in the measuring tube could result in an increase in measuring errors.
Avoid
the following locations:
• At the highest point of a pipeline. Risk of air accumulating.
• Directly upstream from a free pipe outlet in a vertical pipeline.
A0003202
Installing pumps
Do not install the sensor on the intake side of a pump. This precaution is to avoid low pressure and the conse-
quent risk of damage to the lining of the measuring tube. Information on the lining's resistance to partial
vacuum
→
ä
21.
It might be necessary to install pulse dampers in systems incorporating reciprocating, diaphragm or peristaltic
pumps. Information on the measuring system's resistance to vibration and shock
→
ä
19.
A0003203
h
2 x DN