Special cases
Metal, ungrounded piping
In order to prevent outside influences on measurement, it is necessary to use ground cables to connect each
sensor flange to its corresponding pipe flange and ground the flanges. Connect the transmitter or sensor con-
nection housing, as applicable, to ground potential by means of the ground terminal provided for the purpose
(see diagram).
The ground cable for flange-to-flange connections can be ordered separately as an accessory from
Hauser
→
ä
45.
• DN
≤
300 (12"): The ground cable is in direct connection with the conductive flange coating and is secured
by the flange screws (A).
• DN
≥
350 (14"): The ground cable connects directly to the metal transport bracket (B).
"
Caution!
Also, observe company-internal grounding guidelines.
A0004376
Potential equalization with equalizing currents in ungrounded, metal pipes
(ground cable: copper wire, at least 6 mm
2
/10 AWG)
A
Installing ground cable at DN
≤
300 (12")
B
Installing ground cable at DN
≥
350 (14")
Pre-installed ground cable for DN
≤
300 (12") (order option)
Ground cables which are preinstalled on the sensor flange, are also available. These ground cables can be
mounted and connected electrically to the piping in different ways:
• Using a screw on the side of the pipe flange (a)
• Using the flange screws (b)
• Using a pipe clip installed around the pipe (c)
A006117
Possibilities for connecting and mounting pre-installed ground cables
(ground cable: copper wire at least 6 mm
2
/10 AWG)
A
B
a
b
c