Partially filled pipes
Partially filled pipes with gradients necessitate a drain-type configuration. The Empty Pipe Detection function
offers additional protection by detecting empty or partially filled pipes.
"
Caution!
Risk of solids accumulating. Do not install the sensor at the lowest point in the drain. It is advisable to install a
cleaning valve.
A0003204
Down pipes
Install a siphon or a vent valve downstream of the sensor in down pipes h
≥
5 m (16,4 ft). This precaution is
to avoid low pressure and the consequent risk of damage to the lining of the measuring tube. This measure also
prevents the system losing prime, which could cause air inclusions.
Information on the lining's resistance to partial vacuum
→
ä
21.
A0008157
Measures for installation in a down pipe
1
Vent valve
2
Siphon
h
Vertical height of the down pipe regarding the sensor
5 x DN
2 x DN
h
2
1