Trouble-shooting
Micropilot M FMR231 with FOUNDATION Fieldbus
76
Hauser
A155
BAD
Device Failure
BAD
Device Failure
Device needs maintenace now
Electronic Failure
A160
BAD
Device Failure
BAD
Device Failure
Memory Failure/
Device needs maintenance now
Electronic Failure
A164
BAD
Device Failure
BAD
Device Failure
Memory Failure/
Device needs maintenance now
Electronic Failure
A171
BAD
Device Failure
BAD
Device Failure
Memory Failure/
Device needs maintenance now
Electronic Failure
A231
BAD
Device Failure
BAD
Device Failure
Device needs maintenance now
Unspecified Err
A511
Uncertain
configuration error
GOOD
Non specific
Other
Configuration Error
A512
Uncertain
Non specific
GOOD
Non specific
Other
Unspecified Err
W601
Uncertain
configuration error
GOOD
Non specific
Other
Configuration Error
W611
Uncertain
configuration error
GOOD
Non specific
Other
Configuration Error
W621
Uncertain
Non specific
GOOD
Non specific
simulation active
No Error
E641
(Alarm)
BAD
Device Failure
GOOD
Non specific
Device needs maintenance now
Unspecified Err
E641
(Warning)
Uncertain
Non specific
GOOD
Non specific
Device needs maintenance now
Unspecified Err
E651
(Alarm)
BAD
Device Failure
GOOD
Non specific
Other
Unspecified Err
E651
(Warning)
Uncertain
Non specific
GOOD
Non specific
Other
Unspecified Err
A671
BAD
Device Failure
GOOD
Non specific
Configuration Error
No Error
Code
PV Status
SV Status
PV Substatus
SV Substatus
TV Status
TV Substatus
BLOCK_ER
XD_ERROR