Commissioning
Micropilot M FMR231 with FOUNDATION Fieldbus
50
Hauser
Function "full calibr." (006)
This function is used to enter the distance from the minimum level to the maximum level (=span).
L00-FMR2xxxx-14-00-06-en-009
!
Note!
If
bypass
or
stilling well
was selected in the "
tank shape
"
(002)
function, the pipe diameter is
requested in the following step.
In principle, it is possible to measure up to the tip of the antenna. However, due to considerations
regarding corrosion and build-up, the end of the measuring range should not be chosen any closer
than 50 mm (2”) to the tip of the antenna.
⇒