Micropilot M FMR231 with FOUNDATION Fieldbus
Commissioning
Hauser
53
dist. too big
• This error cannot be remedied by interference echo mapping
• Check the application parameters
(002), (003), (004)
and "
empty calibr.
"
(005)
dist. unknown
If the actual distance is not known, no mapping can be carried out.
manual
A mapping is also possible by manual entry of the range to be suppressed. This entry is made in the
"
range of mapping
"
(052)
function.
"
Caution!
The range of mapping must end 0.5 m (20") before the echo of the actual level. For an empty tank,
do not enter E, but E – 0.5 m (20").
If a mapping already exists, it is overwriten up to the distance specified in
"
range of mapping
"
(052)
. Beyond this value the existing mapping remains unchanged.
Function "range of mapping" (052)
This function displays the suggested range of mapping. The reference point is always the reference
point of the measurement (). This value can be edited by the operator.
For manual mapping, the default value is 0 m.
Function "start mapping" (053)
This function is used to start the interference echo mapping up to the distance given in "
range of
mapping
"
(052)
.
Selection:
• off
→
no mapping is carried out
• on
→
mapping is started
During the mapping process the message "
record mapping
" is displayed.
"
Caution!
A mapping will be recorded only, if the device is not in alarm-state.
⇒
E HAUSER
E
+
–
⇒
E HAUSER
E
+
–
Содержание Micropilot M FMR231
Страница 95: ...Micropilot M FMR231 with FOUNDATION Fieldbus Technical data Endress Hauser 95 ...
Страница 112: ...Micropilot M FMR231 with FOUNDATION Fieldbus Index 112 Endress Hauser ...
Страница 113: ......
Страница 114: ...www endress com worldwide BA229F 00 ae 06 07 52009950 CCS FM SGML 6 0 ProMoDo 52008600 ...