Commissioning
Micropilot M FMR231 with FOUNDATION Fieldbus
54
Hauser
display (008)
The
distance
measured from the reference point to the product surface and the
level
calculated
with the aid of the empty adjustment are displayed. Check whether the values correspond to the
actual level or the actual distance. The following cases can occur:
• Distance correct
−
level correct
→
continue with the next function, "
check distance
"
(051)
• Distance correct
−
level incorrect
→
Check "
empty calibr.
"
(005)
• Distance incorrect
−
level incorrect
→
continue with the next function, "
check distance
"
(051)
⇒
E HAUSER
E
+
–
⇒
⇓
After 3 s, the following message appears
E HAUSER
E
+
–