Micropilot M FMR231 with FOUNDATION Fieldbus
Trouble-shooting
Hauser
73
9.2
System error messages
9.2.1
Current error
Errors which the Micropilot M detects during commissioning or operation are displayed:
•
VU331:
error symbol in the
"measured value" (000)
function
•
VU331 or Hauser operating program:
in the
"diagnostics" (0A)
function group in the
"present error" (0A0)
function
Only the highest priority error is displayed; in the case of multiple errors, you can scroll between
the different error messages by pressing
O
or
S
.
•
FOUNDATION Fieldbus:
– by the status of the main value in the cyclic data telegram
– Diagnostic Block, parameter PARACTUALERROR (present error)
FMR2xx-FF-Diagnosticblock-in-NIConfig
9.2.2
Last error
The last error is displayed in the
"diagnostics" (0A)
function group in the
"previous error"
(0A1)
function. This display can be deleted in the
"clear last error" (0A2)
function.
Содержание Micropilot M FMR231
Страница 95: ...Micropilot M FMR231 with FOUNDATION Fieldbus Technical data Endress Hauser 95 ...
Страница 112: ...Micropilot M FMR231 with FOUNDATION Fieldbus Index 112 Endress Hauser ...
Страница 113: ......
Страница 114: ...www endress com worldwide BA229F 00 ae 06 07 52009950 CCS FM SGML 6 0 ProMoDo 52008600 ...