Installation
Memosens Wave CAS80E
12
Hauser
Ø 40 (1.6)
4
9
0
.5
(1
9
.3
)
1
6
0
.5
(6
.3
)
3
3
0
(1
3
)
50 (1.97)
80.9 (3.2)
2.6 (0.1)
20.5 (0.8)
36 (1.4)
A0041368
5
Dimensions of spectrometer with
50 mm (1.97 in) gap. Dimensions: mm (in)
5.1.2
Installation instructions
• Choose a mounting location that can be easily accessed at a later stage.
• Align the spectrometer so that the measurement gap is rinsed by the flow of medium.
• Do not install the device in places where air pockets and foam bubbles form.
• Ensure that upright posts and assemblies are fully secured and vibration-free.
To ensure correct measurement, the windows at the measurement gap must be free from
any sedimentation. The best way to ensure this is through the use of a cleaning unit
(accessory) operated by compressed air.
‣
In the event of horizontal orientation, mount the spectrometer in such a way that air
bubbles can escape from the measurement gap (do not point it downwards).
5.1.3
Orientation
A0013268
6
Alignment, arrow points in
the flow direction
‣
Align the spectrometer in such a way that the
measurement gap is rinsed with the flow of medium
and air bubbles are removed.