Memosens Wave CAS80E
Installation
Hauser
19
1
3
5
6
2
7
2
2
4
A0013361
16
Connection diagram with bypass
1
Main pipe
2
Manually actuated or solenoid valves
3
Medium outlet
4
Medium return
5
Flow assembly
6
Medium inflow
7
Medium sampling
‣
Connect the medium inflow and outlet to the hose connections of the assembly
→ 16, 19.
This fills the assembly from below and ensures that the assembly is self-venting.
• The flow rate must be at least 100 ml/h (0.026 gal/h).
• Take the extended response times into consideration.
2
1
3
4
A0013434
17
Connection diagram with open outlet, arrow points in the flow direction
1
Pump
2
Flow assembly
3
Open outlet
4
Filter unit
As an alternative to operation in the bypass, it is also possible to direct the sample flow
from a filter unit with an open outlet through the assembly → 17, 19.