Memosens Wave CAS80E
Installation
Hauser
15
5.2.3
Immersion assembly
A0013267
11
CYA112 immersion assembly and CYH112
holder installed horizontally, fixed installation
The installation angle is 90°.
‣
Align the spectrometer in such a way that
the measurement gap is rinsed with the
flow of medium and air bubbles are
removed.
A0013270
12
CYA112 immersion assembly and CYH112
holder installed vertically, suspended from a
chain
The installation angle is 0°.
‣
Ensure that the spectrometer is
adequately cleaned. There must be no
buildup on the optical windows.
For detailed information on installing the immersion assembly and the holder, see
BA00432C and BA00430C
5.2.4
Flow assembly 71110000
Mount the flow assembly only on a spectrometer with a gap of 2 mm (0.08 in) or
10 mm (0.4 in).
1
2
3
4
A0013352
13
Spectrometer with flow assembly
1
Transmitter
2
Memosens Wave CAS80E
3
Flow assembly
4
Holder