Memosens Wave CAS80E
Operation
Hauser
27
Example:
This type of adjustment can be selected if the measured values are compared to the
laboratory values over a longer period of time and all values are too low by a constant
factor, e.g. 10%, in relation to the laboratory value (target sample value).
In the example, the adjustment is made by entering the factor 1.1.
1
2
3
4
5
6
7
8
1
2
3
4
5
6
x
y
f = 1.1
a
b
A0039329
23
Principle of factor calibration
x
Measured value
y
Target sample value
a
Factory calibration
b
Factor calibration
1-point calibration
The measured error between the measured value of the device and the laboratory
measured value is too large. This is corrected by a 1-point calibration.
1
2
3
4
5
6
7
8
1
2
3
4
5
6
x
y
a
b
A0039320
24
Principle of a 1-point calibration
x
Measured value
y
Target sample value
a
Factory calibration
b
Application calibration