Memosens Wave CAS80E
Technical data
Hauser
39
Measured variable
Drift over 100 days in % of end of measuring range
SAC
0.02
CODeq
0.01
TOCeq
0.004
BODeq
0.0015
Nitrate NO3-N
0.002
APHA Hazen true
0.01
APHA Hazen apparent
0.01
Drinking water
Measured variable
Drift over 100 days in % of end of measuring range
Turbidity
0.02
TSS
0.02
SAC
0.02
TOCeq
0.002
Nitrate NO3-N
0.002
APHA Hazen true
0.01
APHA Hazen apparent
0.01
Surface water
Measured variable
Drift over 100 days in % of end of measuring range
Turbidity
0.02
TSS
0.02
SAC
0.02
CODeq
0.005
BODeq
0.0008
Nitrate NO3-N
0.002
Detection limit
The detection limits were determined for the individual measured variables in ultrapure
water under laboratory conditions based on DIN ISO 15839.
Wastewater treatment plant inlet
Measured
variable
2 mm (0.08 in) gap
10 mm (0.4 in) gap
50 mm (1.97 in) gap
TSS
20 mg/l
4 mg/l
0.8 mg/l
SAC
1 1/m
0.2 1/m
0.04 1/m
CODeq
10 mg/l
2 mg/l
0.4 mg/l
BODeq
2.5 mg/l
0.5 mg/l
0.1 mg/l
Wastewater treatment plant outlet
Measured
variable
2 mm (0.08 in) gap
10 mm (0.4 in) gap
50 mm (1.97 in) gap
Turbidity
12.5 FAU
2.5 FAU
0.5 FAU
TSS
12.5 mg/l
2.5 mg/l
0.5 mg/l