General troubleshooting
When troubleshooting, the entire measuring point must be taken into account:
• Transmitter
• Electrical connections and cables
• Assembly
• Spectrometer
The possible causes of error in the following table relate mainly to the spectrometer.
Problem
Check
Solution
Nothing displayed, no
reaction from the
spectrometer
• Power supplied to transmitter?
• Spectrometer connected correctly?
• Buildup on optical windows?
‣
Connect mains voltage.
‣
Establish correct connection.
‣
Clean the spectrometer.
Display value too high or
too low
• Buildup on optical windows?
• Spectrometer calibrated?
‣
Clean the windows.
‣
Calibrate the spectrometer.
Display value fluctuating
greatly
• Air bubbles in the measurement
gap?
• Is the mounting location correct?
‣
Clean the windows.
‣
Select a different mounting location.
‣
Adjust measured value filter.
Measured value drift
Buildup on optical windows?
‣
First clean the spectrometer.
‣
Record the reference spectrum.