Installation
Memosens Wave CAS80E
20
Hauser
5.2.5
Cleaning unit
L
CAUTION
Residual medium and high temperatures
Risk of injury!
‣
When working with parts that are in contact with the medium, protect against residual
medium and elevated temperatures.
‣
Wear protective goggles and safety gloves.
Preparatory steps:
1. Mount the compressed air cleaning on the spectrometer before installing in the
measuring point.
2. Remove the spectrometer from the medium if the device is already in the process.
3. Clean the spectrometer.
Spectrometer with 2 mm (0.08 in) or 10 mm (0.4 in) gap:
1.
Insert the elbow plug into the mounting borehole behind the measurement gap as far
as the end stop (hand-tight).
2. Screw the elbow plug tight.
3.
Connect the hose of the compressed air supply at the installation location to the
opening of the elbow plug.
4. Use the hose piece with hose coupling provided with the sensor if desired.
Spectrometer with 50 mm (2 in) gap:
1.
Insert the air distributor into the mounting boreholes behind the measurement gap
as far as the end stop (hand-tight).