Memosens Wave CAS80E
Maintenance
Hauser
33
Type of fouling
Cleaning measure
Lime deposits
‣
Immerse the spectrometer in 1-5% hydrochloric acid (for a few minutes).
Deposit buildup on
the optics
There may be deposit buildup in the non-visible range (UV). Therefore, always clean
the optics.
‣
Rinse the spectrometer with copious amounts of water.
‣
Wet a lint-free cloth with 5-10% phosphoric acid or 5-10% hydrochloric acid.
‣
Insert the cloth into the measurement gap and leave for a maximum of 10 minutes.
‣
Move the cloth back and forth to remove dislodged dirt particles.
‣
Wet the brush supplied with acid.
‣
Use the brush to clean the windows.
After cleaning:
‣
Rinse the spectrometer with copious amounts of water.