Activation code
You require activation codes for:
• Additional functions,e.g. fieldbus communication
• Software upgrades
If activation codes are available for your device, these codes are provided on the inner
nameplate. The corresponding device functions are activated at the factory. You only
require the codes if servicing the device.
1. Enter the activation code: Menu/Setup/General settings/Extended setup/Data
management/Activation code.
2. Confirm your entry.
Your new hardware or software function is activated and can be configured.
The table below tells you what functions an activation code enables:
Function
Activation code beginning with
Second Memosens input
062...
Two current outputs (BASE-E module only)
081...
HART
0B1...
PROFIBUS DP
0B3...
Modbus TCP
0B4...
Modbus RS485
0B5...
EtherNet/IP
0B6...
Measuring range switching, set 1
211...
Measuring range switching, set 2
1)
1)
You receive two activation codes when you order the device with the "Measuring range switching" option. Enter
both in order to have two sets for measuring range switching.
212...
Feedforward control
220...
Chemoclean Plus
25...