Operation
Liquistation CSF48
164
Hauser
10.6 Additional functions
10.6.1 Limit switches
There are different ways of configuring a limit switch:
• Assigning a switch-on and switch-off point
• Assigning a switch-on and switch-off delay for a relay
• Setting an alarm threshold and also outputting an error message
• Starting a cleaning function
Path: Menu/Setup/Additional functions/Limit switches/Limit switches 1 to 8
Function
Options
Info
Source of data
Options
• None
• Connected sensors
• Connected inputs
• Temperature sensors
Factory setting
None
Specify the input or output which should be the source of
data for the limit switch.
The sources of data on offer depend on your device
version. All the sensors and controllers connected to
inputs are available for selection.
Measuring value
Options
• None
• Depends on theSource of
data
Factory setting
None
The measured value you can select depends on the
option selected under "Source of data".
Measured valuedepending on theSource of data
Source of data
Measured value
pH Glass
Options
• Raw value mV
• pH
• Temperature
pH ISFET
ORP
Options
• Temperature
• ORP mV
• ORP %
Oxygen (amp.)
Options
• Temperature
• Partial pressure
• Concentration liquid
• Saturation
• Raw value nA(only Oxygen (amp.))
• Raw value
s(only Oxygen (opt. WW))
Oxygen (opt. WW)
Содержание Liquistation CSF48
Страница 2: ......
Страница 39: ...Liquistation CSF48 Electrical connection Endress Hauser 39 a0016537 Fig 41 E basic module wiring diagram ...
Страница 267: ...Liquistation CSF48 Endress Hauser ...
Страница 268: ...www addresses endress com ...