![Endress+Hauser Liquistation CSF48 Скачать руководство пользователя страница 118](http://html.mh-extra.com/html/endress-hauser/liquistation-csf48/liquistation-csf48_operating-instructions-manual_2407297118.webp)
Operation
Liquistation CSF48
118
Hauser
Start condition
Options
• Immediate
• Date/time
Factory setting
Immediate
The sampling program can be started either
immediately or at a specific, configurable time.
If Start condition
Immediate
is selected:
Sample at start
Options
• Yes
• No
Factory setting
Yes
Yes:
The first sample is taken when the program is
started.
No:
The system waits for the interval to elapse before the
first sample is taken.
If Start condition
Date/time
is selected:
Start date
01.01.2000 to 31.12.2099
Factory setting
DD.MM.YYYY
Set the start date for the sampling program. The
format depends on the option configured under
general settings.
Start time
00:00:00 to 23:59:59
Factory setting
HH:MM:SS (24h)
Set the time when the sampling program is started.
The format depends on the option configured under
general settings.
Stop condition
Options
• Program end
• Continuous
Factory setting
Program end
Program end:
The device stops sampling automatically once it has
run through the set program.
Continuous:
The device runs through the set program
continuously in an infinite loop. Do not forget to
empty the bottles.
Assignment bin. output
Options
• No binary output configured
• Binary output S:x
Factory setting
No binary output configured
Assignment of the binary output to the program
cycle.
Inputs
Settings for the inputs can be made as described in
the "Inputs" section.
Path: Menu/Setup/Sampling programs/Setup program/New/Basic
Function
Options
Info
Содержание Liquistation CSF48
Страница 2: ......
Страница 39: ...Liquistation CSF48 Electrical connection Endress Hauser 39 a0016537 Fig 41 E basic module wiring diagram ...
Страница 267: ...Liquistation CSF48 Endress Hauser ...