Liquistation CSF48
Installation
Hauser
27
5.3
Sampling with a flow assembly
The sample is taken directly from the flow assembly which is installed in the sampler stand.
The flow assembly is used when sampling in pressurized systems, e.g.:
• Containers located at a higher level
• Pressurized pipes
• Pumping with external pumps
The flow rate should be 1000 to 1500 l/h.
The outlet of the flow assembly must be unpressurized (e.g. drain, open channel).
a0013127
Fig. 19: Flow assembly
1
2
3
Flow assembly inlet: ¾"
Sample connection
Flow assembly outlet: 1¼"