9.2.3
Display settings
Path: Menu/Display/Operation
Function
Options
Info
Contrast
5 to 95 %
Factory setting
50 %
Adjust the screen settings to suit your working
environment.
Backlight="Automatic"
The backlighting is switched off automatically after a
short time if a button is not pressed. It switches back on
again as soon as you press the navigator button.
Backlight="On"
The backlighting does not switch off automatically.
Backlight
Options
• On
• Off
• Automatic
Factory setting
Automatic
Screensaver
Options
• Off
• Automatic
Factory setting
Off
The screensaver switches off the display if more than 5
minutes have elapsed since the last action executed in
the software. The device continues to operate as normal,
even when the display is switched off.
Press the Navigator button to reactivate the display.
Screen rotation
Options
• Manual
• Automatic
Factory setting
Manual
If "Automatic" is selected, the single-channel measured
value display switches from one channel to the next
every second.
Current program:
Read only
The name of the sampling program currently selected is
displayed.
Status
Read only
Display "Active":
The sampling program has been started and the device
takes a sample as per the set parameters.
Display "Inactive":
No sampling program has been started, or a program
that was running has been stopped.
Start
Action
The selected sampling program is started.
User definable screens
Meas. screen 1
...
Meas. screen 6
You can create 6 measuring screens of your own and
give them a name.
As the functions are the same for all 6 measuring
screens, only one measuring screen is described below.
Meas. screen
Options
• Off
• On
Factory setting
Off
Once you have defined your own measuring screen, you
can switch it on here. You can find the new screen under
"User definable screens" in the "All measured values"
measuring mode.
Label
Customized text,
20 characters
Name of the measuring screen
Appears in the status line of the measuring display.
Содержание Liquistation CSF48
Страница 2: ......
Страница 39: ...Liquistation CSF48 Electrical connection Endress Hauser 39 a0016537 Fig 41 E basic module wiring diagram ...
Страница 267: ...Liquistation CSF48 Endress Hauser ...
Страница 268: ...www addresses endress com ...