Liquistation CSF48
Diagnostics and troubleshooting
Hauser
211
Overcurrent
Read only
No: no error
Yes: the heating is defective
-> Contact the Service Department
Overcurrent
Read only
No: no error
Yes: the fan is defective
-> Contact the Service Department
Sample compartment
Read only
The
current
temperature of the sample compartment is
displayed.
Sample compartment
Read only
When you start the heating test, the temperature at the
start time is displayed
Heating test offor Heating test on -> progress of test is displayed
Start Test
Action
Start the heating test.
Stop test
Action
End the heating test.
Manual sampling
Bottle configuration
Read only
Bottle volume
Read only
Distributor position
Options
• Front
• Bottle 1
...
• Back
Select which bottle should be filled with the sample.
Sample volume
50 to 2000 ml
Factory setting
100 ml
You can change the sample volume in the version with
the peristaltic pump.
Sample volume
Factory setting
200 ml
The sample volume is preset at the factory in the version
with the vacuum pump.
Start sampling
Action
Peristaltic pump(only for version with peristaltic pump)
Pump purge
Action
Pump purge, to stop
press ESC
Read only
Current pump run time
Read only
Power supply
Read only
The current supply voltage is displayed.
With AC power supply: 24 V ±0.5 V
With DC power supply: 22 to 28 V
Motor current
Read only
The current consumption of the pump is displayed.
Diagnostics/Systemtest/Reset
Function
Options
Info