Diagnostics and troubleshooting
Liquistation CSF48
210
Hauser
11.12 Resetting the measuring device
Main value
Depends on the sensor
Factory setting
Depends on the sensor
Set the desired simulation value.
Sim. temperature
Options
• On
• Off
Factory setting
Off
If you simulate the temperature measured value, this is
indicated on the display by a simulation icon in front of
the temperature.
Temperature
-50.0 to +250.0
˚
C
(-58.0 to 482.0
˚
F)
Factory setting
20.0
˚
C (68.0
˚
F)
Set the desired simulation value.
Diagnostics/Systemtest/Reset
Function
Options
Info
Power supply
Read only
The current supply voltage is displayed.
Cooling system(only for version with sample compartment temperature regulation)
Check cooling
Power supply
Read only
The current supply voltage is displayed.
With AC power supply: 24 V ±0.5 V
With DC power supply: 22 to 28 V
Overcurrent
Read only
No: no error
Yes: the fan in the climate control module is defective
-> Contact the Service Department
Sample compartment
Read only
The
current
temperature of the sample compartment is
displayed.
Sample compartment
Read only
When you start the cooling test, the temperature at the
start time is displayed
Cooling test offor Cooling test on -> progress of test is displayed
Start Test
Action
Start the cooling test.
Stop test
Action
End the cooling test.
Check heating
Power supply
Read only
The current supply voltage is displayed.
With AC power supply: 24 V ±0.5 V
With DC power supply: 22 to 28 V
Diagnostics/Simulation
Function
Options
Info