Liquistation CSF34
Operation
Hauser
99
Menu/Setup/Sampling programs/Setup program/New/Basic
Function
Options
Info
Stop condition
Selection
• Program end
• Continuous
Factory setting
Program end
Program end
The device stops sampling automatically
once it has run through the set program.
Continuous
The device runs through the set program
continuously in an infinite loop. Do not
forget to empty the bottles.
Assignment bin. output
Selection
• No binary output
configured
•
Binary output
Sx
Factory setting
No binary output
configured
Assignment of the binary output to the
program cycle.
Inputs
Settings for the inputs can be made as
described in the "Inputs" section.
10.3.3 Program types: Standard and Advanced
Standard program:
Comprises a maximum of five subprograms
Advanced program:
• Comprises a maximum of 24 subprograms.
• These subprograms can be run simultaneously or consecutively.
• Each event subprogram can consist of up to 3 conditions.
• As the device contains dual bottle trays, you can assign a program easily, and easily
detect a change in the program.
Settings for the Standard program
Menu/Setup/Sampling programs
Function
Options
Info
Setup program
New
A list of all the programs created is
displayed. For this reason, it is often helpful
to add an "S" for Standard in the program
name.
Standard
Program name
Free text
Use a distinct name for your sampling
program. The program name can be up to 16
characters long.
Bottle volume
0 to 100000 ml
0 to 20000 ml
Factory setting
• 30000 ml
• 20000 ml
Set the bottle volume. The preset value
depends on the bottle configuration
configured. The bottle volume is always 30 l
for individual containers. The preset value
depends on the bottle configuration. The
bottle volume is always 20 l for individual
containers.
In the case of asymmetric distribution, e.g. 6
x 3 l + 2 x 13 l, you can set the bottle volume
on the left and right in the menu items that
follow.
Bottle configuration
Choice of all possible bottle
combinations
The ordered bottle configuration is preset or
the configuration selected in the setup is
displayed.