10.3.1 Overview of the possible program types
The graphic below explains the various ways sampling can be controlled on the basis of a
flow curve:
a
b
c
d
e
A0014045
66
Sampling control
Q
Flow
t
Time
a. Flow curve
b.
Time-proportional sampling (CTCV)
A constant sample volume (e.g. 50 ml) is taken at regular
intervals (e.g. every 5 min).
c.
Volume-proportional sampling (VTCV)
A constant sample volume is taken at variable intervals
(depending on the inflow volume).
d.
Flow-proportional sampling (CTVV)
A variable sample volume (the sample volume depends on the
flow rate) is taken at regular intervals (e.g. every 10 min).
e.
Event-controlled sampling
Sampling is triggered by an event (e.g. pH limit value). Sampling
can be time-paced, volume-paced or flow-paced, or single
samples can be taken.