Mounting
Liquistation CSF34
22
Hauser
Application example: Taking samples from pressure piping
A0023437
15
Taking samples from pressure piping
V1 Diaphragm valve
V2 Ball valve
3
Flow assembly
Use diaphragm valve 1 to set the flow rate to a
maximum of 1000 l/h to 1500 l/h. When the
sampling cycle begins, one of the relay outputs can
be used to control and open ball valve 2. The
medium flows through the pipe and the flow
assembly and into the outflow. Once an adjustable
delay time has elapsed, the sample is taken directly
from the flow assembly. Ball valve 2 is closed again
once the sample has been taken.
The ball valve and the diaphragm valve are not
included in the scope of supply. If necessary,
please request a quote from your Endress
+Hauser sales center.
5.4
Post-installation check
1. Verify that the suction line is securely connected to the device.
2. Visually check that the suction line is installed correctly from the sampling point to
the device.
3. Verify that the distribution arm is correctly engaged.
4. Allow the sampler to rest for at least 12 hours after setup and before switching it on.
Otherwise you may cause damage to the climate control module.