Liquistation CSF34
Technical data
Hauser
201
15.3 Protocol-specific data
HART
Manufacturer ID
11
h
Device type
119D
h
Device revision
001
h
Device description files (DD/DTM)
www.endress.com/hart
Device Integration Manager DIM
Device variables
Supported features
PDM DD, AMS DD, DTM,
PROFIBUS DP
Manufacturer ID
11
h
Device type
155C
h
Profile version
3.02
Device database files (GSD files)
www.endress.com/profibus
Device Integration Manager DIM
Output variables
Supported features
• 1 MSCY0 connection (cyclical communication, master class 1
to slave)
• 1 MSAC1 connection (acyclical communication, master class 1
to slave)
• 2 MSAC2 connections (acyclical communication, master class
2 to slave)
• Addressing using DIL switches or software
• GSD, PDM DD, DTM
Modbus RS485
Protocol
RTU/ASCII
Function codes
03, 04, 06, 08, 16, 23
Broadcast support for function codes
06, 16, 23
Output data
16 measured values (value, unit, status), 8 digital values (value,
status)
Input data
4 setpoints (value, unit, status), 8 digital values (value, status),
diagnostic information
Supported features
Address can be configured using switch or software
Modbus TCP
TCP port
502
TCP connections
3
Protocol
TCP
Function codes
03, 04, 06, 08, 16, 23
Broadcast support for function codes
06, 16, 23
Output data
16 measured values (value, unit, status), 8 digital values (value,
status)
Input data
4 setpoints (value, unit, status), 8 digital values (value, status),
diagnostic information
Supported features
Address can be configured using DHCP or software