
Operation
Liquistation CSF34
94
Hauser
Menu/Setup/Sampling programs/Setup program/New/Basic
Function
Options
Info
Stop condition
Selection
• Program end
• Continuous
Factory setting
Program end
Program end
The device stops sampling automatically
once it has run through the set program.
Continuous
The device runs through the set program
continuously in an infinite loop. Do not
forget to empty the bottles.
Assignment bin. output
Selection
• No binary output
configured
•
Binary output
Sx
Factory setting
No binary output
configured
Assignment of the binary output to the
program cycle.
Inputs
Settings for the inputs can be made as
described in the "Inputs" section.
Settings with the Basic program type with several bottles
Sampling mode
=
Time/flow paced CTVV
Menu/Setup/Sampling programs/Setup program/New/Basic
Function
Options
Info
Sampling volume input
Selection
• No flow input
configured
• Binary input S:x
• Current input S:x
Factory setting
No flow input configured
Select the sampling volume input. The
binary input or the current input must be
configured for this function. Only the inputs
configured as a sampling volume input are
displayed.
Sampling interval
00:01:00 ... 99:59:00
HH:MM:SS
Factory setting
00:10:00 HH:MM:SS
00:10:00 HH:MM:SS
HH:MM:SS
Factory setting
00:10:00 HH:MM:SS
Set the sampling interval.
If is selected:
Sampling volume input
Binary input
Sampling volume
10 to 10000 ml
Factory setting
20 ml
Set the sampling volume.
The dosing accuracy and the repeatability of
a sample volume < 20 ml can vary,
depending on the specific application.
If is selected:
Sampling volume input
Current input
Sampling volume 20mA
10 to 10000 ml
Factory setting
100 ml
Set what sampling volume should be
grabbed at 20 mA.
The dosing accuracy and the repeatability of
a sample volume < 20 ml can vary,
depending on the specific application.
Flow calculation
Selection
• Current
• Average flow
Factory setting
Current
Current:
The current flow is converted to the
sampling volume at the time of sampling.
Average flow:
The system calculates the mean between the
last and the current sample and sets the
sampling volume accordingly.
Содержание Liquistation CSF34
Страница 2: ......
Страница 4: ...Table of contents Liquistation CSF34 4 Endress Hauser Index 207...
Страница 209: ......
Страница 210: ...www addresses endress com 71588428 71588428...