Liquistation CSF34
Mounting
Hauser
17
5.1.2
Installation site
For version with sample pump
1
2
3
4
A0024411
7
Liquistation mounting conditions
Mounting conditions
Route the suction line with a downward gradient to the sampling point.
Never mount the sampler in a place where it is exposed to aggressive gases.
Avoid siphoning effects in the suction line.
Do not route the suction line with an upward gradient to the sampling point.
Note the following when erecting the device:
• Erect the device on a level surface.
• Connect the device securely to the surface at the fastening points.
• Protect the device against additional heating (e.g. heaters or direct sunlight).
• Protect the device against mechanical vibrations.
• Protect the device against strong magnetic fields.
• Make sure air can circulate freely at the side panels of the cabinet. Do not mount the
device directly against a wall. Allow at least 150 mm (5.9 in.) from the wall to the left
and right.
• Do not erect the device directly above the inlet channel of a wastewater treatment plant.