Technical data
Liquistation CSF34
202
Hauser
EtherNet/IP
Log
EtherNet/IP
ODVA certification
Yes
Device profile
Generic device (product type: 0x2B)
Manufacturer ID
0x049E
h
Device type ID
0x109
Polarity
Auto-MIDI-X
Connections
CIP
12
I/O
6
Explicit message 6
Multicast
3 consumers
Minimum RPI
100 ms (default)
Maximum RPI
10000 ms
System integration
EtherNet/IP
EDS
Rockwell
Add-on-Profile Level 3, Faceplate for Factory
Talk SE
IO data
Input (T → O)
Device status and diagnostic message with
highest priority
Measured values:
• 16 AI (analog input) + Unit
• 8 DI (discrete input) + Status
Output (O → T)
Actuating values:
• 4 A0 (analog output) + unit
• 8 DO (discrete output) + Status
Web server
The web server enables full access to the device configuration, measured values, diagnostic
messages, logbooks and service data via standard WiFi/WLAN/LAN/GSM or 3G routers
with a user-defined IP address.
TCP port
80
Supported features
• Remote-controlled device configuration
• Save/restore device configuration (via SD card)
• Logbook export (file formats: CSV, FDM)
• Access to web server via DTM or Internet Explorer
15.4 Power supply
Supply voltage
100 to 120/200 to 240 V AC ±10 %, 50/60 Hz
The device does not have a power switch.
A fuse with a maximum rating of 10 A must be provided by the customer. Observe the
local regulations for installation.
Power consumption
• Version with vacuum pump: 290 VA
• Version with peristaltic pump: 290 VA
• Version with sampling assembly: 290 VA
• Version with 24V power supply: 240 W
Electrical connection
See the "Electrical connection" section ()