Gammapilot M FMG60 FOUNDATION Fieldbus
Wiring
Hauser
29
Feature 30 of the ordering information: power supply
wiring/output wiring
Connection compartment
with terminals for
Supply volt-
age
Signal out-
put
A
Non-Ex; non-Ex
1
2
1
2
A0018082
B
Ex e; Ex ia
1
2
C
Ex e; Ex e
1
1
D
Ex d (XP); Ex d (XP)
1
1
E
Ex d (XP); Ex ia (IS)
1
2
F
Dust-Ex; Dust-Ex
1
1
G
Ex e, Dust-Ex; Ex e, Dust-Ex
1
1
H
Ex d, Dust-Ex; Ex d, Dust-Ex
1
1
J
Ex e, Dust-Ex; Ex ia, Dust-Ex
1
2
K
Ex d, Dust-Ex; Ex ia, Dust-Ex
1
2
L
Dust-Ex; Ex ia
1
2
5.4
Fieldbus connectors
For the versions with a fieldbus connector M12 or 7/8", the housing does not need to be
opened to connect the signal cable.
5.4.1
Pin assignment for connector M12
2
1
3
4
A0011175
PIN
Meaning
1
2
Not assigned
3
Signal –
4
Ground
5.4.2
Pin assignment for connector 7/8"
4
2
3
1
A0011176
PIN
Meaning
1
Signal –
2
3
Shielding
4
Not assigned
5.5
Foundation Fieldbus cable specifications
Always use a twisted, shielded two-wire cable. The cable specifications can be taken from
the FF Specification or IEC 61158-2. The following are examples of suitable cable types:
Non-hazardous area:
• Siemens 6XV1 830-5BH10 (gray)
• Kerpen CEL-PE/OSCR/PVC/FRLA FB-02YS(ST)YFL (gray)
• Belden 3076F (orange)
Содержание Gammapilot M FMG60 FOUNDATION Fieldbus
Страница 104: ...www addresses endress com 71401076 71401076...