Operation
Gammapilot M FMG60 FOUNDATION Fieldbus
40
Hauser
FieldCare supports the following functions:
• Configuration of transmitters in online mode
• Loading and saving device data (upload/download)
• Documentation of the measuring point
Connection options:
Commubox FXA291 with ToF Adapter FXA291 via service interface
6.4
Locking/unlocking configuration
6.4.1
Software locking
Enter a number unequal to 2457 in the
"Unlock parameter" (*A4)
function in the
"Diagnostics" (*A)
function group. The
symbol appears on the display. No more
entries can be made.
If you try to change a parameter, the device goes to the
"Unlock parameter" (*A4)
function. Enter "2457". Parameters can now be modified again.
6.4.2
Hardware locking
Press , and simultaneously. No more entries can be made. If you try to change a
parameter, the following appears:
Local display
Unlock parameter 0A4
Hardware locked
Press , and simultaneously. The
"Unlock parameter" (*A4)
function appears.
Enter "2457". Parameters can now be modified again.
Hardware locking can only be unlocked again via the display by pressing the ,
and keys simultaneously once again. It is not possible to unlock the hardware by
communication.
6.5
Resetting to the default configuration
It is always advisable to reset the customer parameters if you want to use a device with an
unknown history. Effects of resetting:
• All customer parameters are reset to their default values.
• Linearization is switched to
"Linear"
but the table values are kept. The table can be
switched back on in the
"Linearization" (*4)
function in the
"Linearization" (*40/*46)
function group.
To perform a reset, enter "33333" in the
"Reset" (*A3)
function in the
"Diagnostics" (*A)
function group.
L
CAUTION
‣
A reset may negatively impact the measurement. As a rule, a basic setup must be
performed again after a reset. All the calibration data are deleted after a reset. A
complete calibration is needed to put the measurement back into operation.
The default values of each parameter are shown in bold in the menu overview in the
appendix.