Installation
Gammapilot M FMG60 FOUNDATION Fieldbus
24
Hauser
L
CAUTION
‣
The water must always be introduced from the bottom to ensure that the water cooling
jacket is completely filled.
4.3.2
Required flow
The required flow rate depends on:
• The ambient temperature at the water cooling jacket
• The inlet temperature
• The measuring length of the Gammapilot M
Typical values are given in the following tables:
Ambient temperature T
A
= 75 °C (167 °F)
Inlet temperature
°C (°F)
Measuring length in mm (in)
50 (1.97) 200 (7.87) 400 (15.7) 800 (31.5) 1200 (47.2) 1600 (63) 2000 (78.7)
20 (68)
30 l/h
30 l/h
30 l/h
41 l/h
55 l/h
70 l/h
84 l/h
25 (77)
30 l/h
30 l/h
30 l/h
45 l/h
61 l/h
77 l/h
93 l/h
30 (86)
30 l/h
30 l/h
33 l/h
50 l/h
68 l/h
86 l/h
104 l/h
35 (95)
30 l/h
30 l/h
38 l/h
59 l/h
80 l/h
101 l/h
122 l/h
40 (104)
30 l/h
30 l/h
47 l/h
72 l/h
98 l/h
124 l/h
149 l/h
Ambient temperature T
A
= 100 °C (212 °F)
Inlet temperature
°C (°F)
Measuring length in mm (in)
50 (1.97) 200 (7.87) 400 (15.7) 800 (31.5) 1200 (47.2) 1600 (63) 2000 (78.7)
20 (68)
30 l/h
30 l/h
38 l/h
59 l/h
80 l/h
101 l/h
122 l/h
25 (77)
30 l/h
30 l/h
42 l/h
64 l/h
87 l/h
110 l/h
133 l/h
30 (86)
30 l/h
30 l/h
47 l/h
73 l/h
98 l/h
124 l/h
150 l/h
35 (95)
30 l/h
30 l/h
54 l/h
84 l/h
113 l/h
143 l/h
173 l/h
40 (104)
33 l/h
33 l/h
66 l/h
101 l/h
137 l/h
173 l/h
210 l/h
Ambient temperature T
A
= 120 °C (248 °F)
Inlet temperature
°C (°F)
Measuring length in mm (in)
50 (1.97) 200 (7.87) 400 (15.7) 800 (31.5) 1200 (47.2) 1600 (63) 2000 (78.7)
20 (68)
30 l/h
30 l/h
45 l/h
70 l/h
94 l/h
119 l/h
144 l/h
25 (77)
30 l/h
30 l/h
50 l/h
77 l/h
104 l/h
131 l/h
158 l/h
30 (86)
30 l/h
30 l/h
55 l/h
85 l/h
115 l/h
146 l/h
176 l/h
35 (95)
32 l/h
32 l/h
64 l/h
98 l/h
133 l/h
168 l/h
203 l/h
40 (104)
38 l/h
38 l/h
75 l/h
116 l/h
157 l/h
199 l/h
240 l/h
4.4
Post-installation check
After installing the measuring device, carry out the following checks: