Description of device parameters
Deltabar FMD72
98
Options
• x.x
• x.xx
• x.xxx
• x.xxxx
• x.xxxxx
Factory setting
x.x
Damping switch (164)
Navigation
Setup
®
Damping switch (164)
Write permission
No write permissions. Parameter is read only.
Description
Displays the switch position of DIP switch 2 which is used to switch the damping of the output
signal on and off.
Display
• Off
The output signal is not damped.
• On
The output signal is damped. The attenuation constant is specified in the "Damping value" (017)
(184) parameter
Factory setting
On
Alarm cur.switch (165)
Navigation
Setup
®
Advanced Setup
®
Current output
®
Alarm cur.switch (165)
Write permission
No write permissions. Parameter is read only.
Description
Displays the switching state of DIP switch 3 "SW/Alarm min."
Display
• SW setting
The alarm current has the value defined in "Output Fail Mode" (051).
• Alarm min.
The alarm current is 3.6 mA, regardless of the software setting.
Corrected press. (172)
Navigation
Setup
®
Corrected press. (172)
Diagnosis
®
Measured values
®
Corrected press. (172)
Write permission
No write permissions. Parameter is read only.6
Description
Displays the measured differential pressure after position adjustment.